Catalog Number CS-25142-F |
Device Problem
Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported "the dilator(large) splite" during use.No patient harm was reported.The device was replaced.The patient's condition is reported as fine.
|
|
Event Description
|
It was reported "the dilator(large) splite" during use.No patient harm was reported.The device was replaced.The patient's condition is reported as fine.
|
|
Manufacturer Narrative
|
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
|
Event Description
|
It was reported "the dilator(large) splite" during use.No patient harm was reported.The device was replaced.The patient's condition is reported as fine.
|
|
Manufacturer Narrative
|
Qn#(b)(4).The customer returned one dilator for evaluation.Signs of use in the form of dried biological material were observed on the returned dilator.Visual inspection of the dilator revealed the tip was split and damaged.The tip also contained white coloration, indicating undue force was applied to the dilator tip during insertion.The total length of the dilator measured to be 5.47" which is within specifications of 5.25-5.75" per product drawing.This dilator contains a double-tapered design to aid in insertion.The outer diameter of the distal portion of the dilator measured to be 0.154" which is within specifications of 0.152-0.164" per product drawing.The outer diameter of the proximal portion of the dilator measured to be 0.184" which is within specifications of 0.183-0.186" per product drawing.The inner diameter of the distal tip could not be accurately measured due to the damage observed.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "do not use excessive force when introducing guidewire or tissue dilator as this can lead to vessel perforation, bleeding, or component damage." the customer report of a damaged dilator tip was confirmed through complaint investigation of the returned sample.Visual inspection confirmed the dilator tip was split and torn.The appearance of this damage is consistent with undue force being applied to the dilator during insertion.Based on these circumstances, unintentional use error likely caused or contributed to this event.Teleflex will continue to monitor and trend for complaints of this nature.
|
|
Search Alerts/Recalls
|