Model Number XOM UNK MICDEBBUR |
Device Problems
Break (1069); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/25/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Analysis results were not available as of the date of this report.A supplemental report will be submitted when analysis is complete.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that during op, the device was not turning the bur.The device was used during the cadaveric course.There was no patient impact associated with the event.Service and repair team indicated that broken bur stuck in the collet.
|
|
Manufacturer Narrative
|
H3: product analysis found that visually, the inner shaft was bent until broken 0.50 inches from the distal end of the inner hub when returned.A portion of the inner shaft 1.12 inches in length was also returned.There was deformation in the distal outside diameter of the hub.The outside diameter of the inner hub should be 0.330 ± 0.002 inches and measured 0.330 inches in the undamaged area and up to 0.359 inches in the deformed area which was out of specification.The hub bushing was rough and worn and was flaring out from the hub assembly.Functional testing could not be performed due to the broken state of the device.In the returned condition, there was an out of specification condition that was related to the complaint (due to physical damage).H6:previous codes b21, c21, d16 are no longer valid.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|