Model Number 3116 |
Device Problems
Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/14/2013 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id 4351-35 lot# serial# (b)(4) implanted: (b)(6) 2013 explanted: (b)(6) 2014, product type lead.Product id 4351-35 lot# serial# (b)(4), implanted: (b)(4) explanted: (b)(4) product type lead product id 4351-35 lot# serial# (b)(4) implanted: (b)(6) 2013 explanted: (b)(6) 2014 product type lead.Product id 4351-35 lot# serial# (b)(4), implanted: (b)(6) 2013 explanted: (b)(6) 2014 product type lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(4), ubd: 30-nov-2014, udi#: (b)(4); product id: 4351-35, serial/lot #: (b)(4), ubd: 30-nov-2014, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for gastric stimulation.It was reported that with their first enterra device, the lead wire broke inside of them 3 months after implant.Their system was replaced because the wires broke and was loose on the side.
|
|
Manufacturer Narrative
|
H3: the returned device was subjected to a series of standard tests that include but is not limited to visual inspection, output and telemetry testing, and functional testing.The returned device passed all testing in the laboratory and no anomalies were identified.The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.Analysis identified that there was a broken conductor wire, the wire was stretched; consistent with overstress damage.The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.The lead was returned segmented; however electrical testing of the returned segment(s) determined that continuity was complete and there were no electrical shorts between the circuits medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|