EDWARDS LIFESCIENCES FEM-FLEX II FEMORAL ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number FEMII018A
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Exsanguination (1841)
Event Date 08/23/2022
Event Type  Injury  
Manufacturer Narrative
Additional narratives: these cannulae are used to divert large volumes of blood between the patient circulation and the cardiopulmonary bypass machine during cardiac surgery and, on occasion, ecmo (off-label use).This malfunction/defect would likely require an exchange of the cannula resulting in a temporary interruption of bypass and possibly significant blood loss.The potential for serious injury before or during use is not remote.The device was not returned for evaluation, as it was discarded per customer.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Event Description
It was reported that the vented port on a 18 fr fem-flex arterial cannula came off from a patient who was on the vv-ecmo.The patient subsequently exsanguinated and expired later.Per customer left internal jugular vein and right fem vein were cannulated and the cannula was in the lijv.The patient was on the ecmo for 3 weeks before the incident happened.
 
Manufacturer Narrative
Based on the reported information the cause of the event was intentional off-label/contra-indicated use for greater than 6 hours, with ecmo, and placed in vasculature other than the femoral.The device history record (dhr) was not able to be reviewed as the device serial number was not provided.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEM-FLEX II FEMORAL ARTERIAL CANNULA
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
reginald santos
one edwards way
mailstop anton 6.1
irvine, CA 92614
9492502731
MDR Report Key15440601
MDR Text Key300075222
Report Number2015691-2022-07939
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140208
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberFEMII018A
Device Catalogue NumberFEMII018A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/23/2022
Initial Date FDA Received09/16/2022
Supplement Dates Manufacturer Received10/25/2022
Supplement Dates FDA Received10/26/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
-
-