• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number 100015555
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2022
Event Type  malfunction  
Event Description
During the af procedure, it was not possible to connect the neck patch.The navlink and patches were replaced, but the issue was not resolved.Later upon checking the amplifier port, bent pins were found which caused a communication issue.The procedure was cancelled.
 
Manufacturer Narrative
Additional information: g3, h2, h3 the results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported bent pins and subsequent cancellation remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ SYSTEM VELOCITY AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key15447091
MDR Text Key305605010
Report Number2184149-2022-00223
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05414734210713
UDI-Public05414734210713
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
K160187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100015555
Device Catalogue Number100014514
Device Lot Number5587047
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/05/2022
Initial Date FDA Received09/19/2022
Supplement Dates Manufacturer Received12/01/2022
Supplement Dates FDA Received12/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-