• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLU SELECT SUCTIONAID TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX BLU SELECT SUCTIONAID TRACHEOSTOMY TUBE; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 101/860/080CZ
Device Problems Disconnection (1171); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/01/2022
Event Type  malfunction  
Event Description
It was reported that during the pre-use check, the pilot balloon got detached from the inflation line.No patient injury.No additional information is available for this complaint.
 
Manufacturer Narrative
Device lot number is unknown, no product information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Event Description
Additional information was received 2022-11-02 from manufacturer receiving facility: this event occurred during patient use, not during pre-use checks, as the returned product had residual patient fluids in the suction line.Additional information was requested to verify; however, no further information was received from the customer.
 
Manufacturer Narrative
Other text: manufacturing device history record review was not performed because the results of the complaint investigation is not batch related issue.A product sample was received for evaluation.Visual inspection found the inflation line was detached from pilot balloon.The cause of the reported problem was traced to the manufacturing process.This issue was escalated to an internal capa and occurrence is being monitored via trend analysis.If the occurrence of this issue increases significantly, additional corrective actions will be taken.Additional event information received (updated b5).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLU SELECT SUCTIONAID TRACHEOSTOMY TUBE
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka,
minneapolis, MN 55442
MDR Report Key15455939
MDR Text Key304539275
Report Number3012307300-2022-19782
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076158
UDI-Public15019517076158
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/860/080CZ
Device Catalogue Number101/860/080CZ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/30/2022
Initial Date FDA Received09/20/2022
Supplement Dates Manufacturer Received11/02/2022
Supplement Dates FDA Received12/01/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-