• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLATELET, PLASMA, RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLATELET, PLASMA, RBC SET Back to Search Results
Model Number 82406
Device Problems Contamination of Device Ingredient or Reagent (2901); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/25/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation is in process, a follow-up report will be provided.
 
Event Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.Donor unit #: (b)(4) there was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: further evaluation of this event has determined that the device did not cause or contribute to a death or serious injury, nor is there a likely potential for death or serious injury associated with this event based on additional investigational information.Signals in the run data file (rdf) indicated that the trima device operated as intended by flagging the procedure to verify wbcs.Root cause: a root cause assessment was performed for this complaint.Based on the available information a definitive root cause could not be determined but it is likely due to one or a combination of the possible causes listed below: * rbc spill over * centrifuge stopped * rbc detector calibration error * possible air block * the orientation of the tubing as loaded in the centrifuge hex holder and the compromised structural integrity of the plasma tubing.Effectively, there is a certain orientation in the hex holder that allows for the rbc line to lie on top of the plasma line and, under certain flow conditions, may cause the plasma line to pinch off.This in turn causes higher flow through the lrs chamber, which may lead to elevated levels of wbcs in the platelet product.
 
Event Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.Donor unit #: w051522059102 there was not a transfusion recipient or patient involved at the time of the residual wbc testing, therefore no patient information is reasonably known at the time of the event.The platelet collection set is not available for return because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL PLATELET, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key15456184
MDR Text Key300395856
Report Number1722028-2022-00309
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583824065
UDI-Public05020583824065
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2024
Device Model Number82406
Device Catalogue Number4824062
Device Lot Number2206082142
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/26/2022
Initial Date FDA Received09/20/2022
Supplement Dates Manufacturer Received10/19/2022
Supplement Dates FDA Received11/08/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-