MEDTRONIC MEXICO S. DE R.L. DE CV ACHIEVE MAPPING CATHETER - 20 MM; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
|
Back to Search Results |
|
Model Number 990063-020 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Myocardial Infarction (1969); Thrombosis/Thrombus (4440)
|
Event Date 09/14/2022 |
Event Type
Injury
|
Event Description
|
It was reported that during a cryo ablation procedure, the patient began to experience dizziness after the transseptal puncture and sheath insertion.After one application of cryo ablation, the patient's symptoms persisted.An angiography was performed and revealed that the patient was experiencing a myocardial infarction.The thrombus in the left circumflex branch was removed by the surgeon which resolved the issue.The case was aborted. no further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary: the 990063-030 mapping catheter with lot 222305509 was returned and analyzed.Visual inspection of the mapping catheter showed it was intact with no apparent issues.No kinks were observed along the shaft, pebax or tip/loop section of the catheter.Functional testing was performed.The mapping catheter was connected to a diagnostic computer using a test cable.All the channel pairs were not displaying any noise.Movement and deflection of the catheter distal to the lemo connector did not induce any interference noise.No anomalies were identified.In conclusion, the clinical issue (myocardial infarction) occurred during the procedure.There is no indication of a relation of the adverse event to the performance and malfunction of the product.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|