It was reported that after the unspecified bd pegasus¿ iv catheter was used on the patient, they developed redness, swelling, and pus at the puncture site, which needed to be reported to the doctor.The dressing was changed and the site disinfected according to the doctor's advice.The following information was provided by the initial reporter, translated from chinese: "on (b)(6) 2022, the patient was admitted to the hospital for traumatic arthritis of the right little finger.On (b)(6) 2022, the patient was sent to the surgery for tendon release of the right little finger.The surgery was performed safely and the patient was given intravenous infusion with indwelling intravenous needle and continuous analgesic pump.If the puncture site is red, swollen and painful, the indwelling needle should be removed.If pus is squeezed at the puncture site, report to the doctor in charge, and change the dressing and disinfection at the needle site according to the doctor's advice.".
|
Unknown manufacturer: there are multiple bd locations where this unspecified bd device may have been manufactured.A catalog and lot number could not be confirmed for this incident and without this information we are unable to determine where the device was manufactured.Therefore, bd corporate headquarters in (b)(4) has been listed in and the (b)(4) fda registration number has been used for the manufacture report number.Medical device expiration date: unknown.Initial reporter phone#: unknown initial reporter e-mail: unknown device manufacture date: unknown.Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received. the lot number is unknown, therefore device history record review (dhr) or quality notification review (qn) could not be performed.Examination of the product involved may provide clarification as to the cause for the reported failure.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly. our business team regularly reviews the collected data for identification of emerging trends.
|