• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA PLATELET+SAMPLER, PLASMA, RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA PLATELET+SAMPLER, PLASMA, RBC SET Back to Search Results
Model Number 82446
Device Problems Contamination of Device Ingredient or Reagent (2901); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/22/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation: per the customer, they switched to a new apheresis platelet set at the beginning of the month when validating the trima software upgrade.Investigation is in process, a follow-up report will be provided.
 
Event Description
The customer reported that they had a positive bacterial detection.There was not a transfusion recipient or patient involved at the time of the unit processing, therefore no patient information is reasonably known at the time of the event.The disposable set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: per the customer, they switched to a new apheresis platelet set at the beginning of the month when validating the trima software upgrade.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.A disposable complaint history search was performed for this lot and found no other reports for similar issues on this lot.Investigation is in process, a follow-up report will be provided.
 
Event Description
The customer reported that they had a positive bacterial detection.There was not a transfusion recipient or patient involved at the time of the unit processing, therefore no patient information is reasonably known at the time of the event.The disposable set is not available for return because it was discarded by the customer.
 
Event Description
The customer reported that they had a positive bacterial detection.There was not a transfusion recipient or patient involved at the time of the unit processing, therefore no patient information is reasonably known at the time of the event.The disposable set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
This report is being filed to provide additional information in h.6 and h.10.Investigation: per the customer, they switched to a new apheresis platelet set at the beginning of the month when validating the trima software upgrade.A review of the device history record (dhr) for this unit showed no irregularities during manufacturing that were relevant to this issue.A disposable complaint history search was performed for this lot and found no other reports for similar issues on this lot.Root cause: a root cause assessment was performed for this complaint.Based on the available information a definitive root cause could not be determined but it is likely due to one or a combination of the possible causes listed below: * improper venipuncture technique introduces bacteria at access site resulting in bacterial growth in product bag.* improper venipuncture technique and no blood diversion performed due to operator error resulting in bacterial contamination of product.* inadequate post-processing laboratory practices such as qc sampling or handling techniques.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA PLATELET+SAMPLER, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
Manufacturer (Section G)
TERUMO BCT
10810 w. collins ave
lakewood CO 80215
Manufacturer Contact
scot hilden
10810 w. collins ave
lakewood, CO 80215
MDR Report Key15473697
MDR Text Key305890601
Report Number1722028-2022-00311
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583824461
UDI-Public05020583824461
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2024
Device Model Number82446
Device Catalogue Number5824462
Device Lot Number2205034351
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/22/2022
Initial Date FDA Received09/23/2022
Supplement Dates Manufacturer Received12/13/2022
01/06/2023
Supplement Dates FDA Received12/27/2022
01/12/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/02/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-