Model Number SENSH1828W
Device Problems Difficult to Remove (1528); Material Deformation (2976); Material Integrity Problem (2978)
Patient Problem Insufficient Information (4580)
Event Date 09/16/2022
Event Type  Injury  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A sentrant sheath was used in an unknown endovascular procedure.  it was reported that during the procedure, there was no resistance advancing the sheath.It was said the patients vascular access was good, straight and the lumen was big enough for the sheath.The patient had also been administered a vasodilator.When the physician tried to withdraw the sheath, it was difficult , when more attempts were made to withdraw it, the sheath broke.The broken sheath was in the patients femoral artery, and the patient required intervention to remove it, and a stent was implanted over the area.  per the physician the cause of removal difficulties was a vascular spasm but because the patient had been administered a vasodilator, it was thought there may be some issue with the sheath such as the hydrophilic coating.  no additional clinical sequalae were provided and the patient is fine.
 
Manufacturer Narrative
B5; additional information recieved ; it was reported the sentrant sheath was used as part of a transcatheter aortic valve replacement(tavr) procedure.It was confirmed the physician could not remove the broken piece of sheath and it is still in the patients femoral artery.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product analysis conclusion; the sheath returned with the dilator partially loaded.Excessive stretching was evident to the sheath when compared to a sheath of the same size.The sheath tip had detached, and the braid wire was unravelled and protruding at the break site.The sheath material was jagged and uneven at the break site.The reported deformation was confirmed through analysis of the returned sheath.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTRANT INTRODUCER SHEATH
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway,ie
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15474995
MDR Text Key300556287
Report Number9612164-2022-03520
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K123990
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/07/2023
Device Model NumberSENSH1828W
Device Catalogue NumberSENSH1828W
Device Lot Number00169259
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/16/2022
Initial Date FDA Received09/23/2022
Supplement Dates Manufacturer Received09/27/2022
11/17/2022
Supplement Dates FDA Received10/18/2022
12/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age87 YR
Patient SexFemale
-
-