• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION HEIDBRINK ELEVATOR #1 STRAIGHT; ELEVATOR, SURGICAL, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION HEIDBRINK ELEVATOR #1 STRAIGHT; ELEVATOR, SURGICAL, DENTAL Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Zimmer biomet complaint: (b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the tip of the instrument fractured during a procedure.There were no consequences or impact to the patient.Another instrument was used to complete the procedure.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6; h10.A visual inspection was conducted on the returned elevator.The elevator shows signs of multiple uses including marking/ scratching on the elevator surface.Medical records were not provided.Review of the dhr identified no deviations or anomalies.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEIDBRINK ELEVATOR #1 STRAIGHT
Type of Device
ELEVATOR, SURGICAL, DENTAL
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15475591
MDR Text Key304788704
Report Number0001032347-2022-00309
Device Sequence Number1
Product Code EMJ
UDI-Device Identifier00841036028862
UDI-Public(01)00841036028862(10)060415E15
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number09-0341
Device Lot NumberE15
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/29/2022
Initial Date FDA Received09/24/2022
Supplement Dates Manufacturer Received10/21/2022
Supplement Dates FDA Received10/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-