Model Number 0998-00-3023-53
Device Problem Circuit Failure (1089)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/11/2022
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) units screen went fuzzy and the metrics could not be made out.The balloon was removed and tagged for biomed.
 
Manufacturer Narrative
It was reported that the customer called for assistance troubleshooting a screen issue on a cs300 during use.It was reported that the screen suddenly went ¿fuzzy¿ and they could not make out any of the metrics.They turned the monitor back and forth and the issue resolved.The fse suggested changing out the console.Customer informed that they were scheduled to take the balloon out anyway so they would leave it as is for now.Text update several minutes later said the balloon was removed successfully and pump tagged for biomed.There are no information regarding the fse who will be assigned/going to that facility to evaluate machine.User resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15479716
MDR Text Key306266532
Report Number2249723-2022-02443
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/11/2022
Initial Date FDA Received09/24/2022
Supplement Dates Manufacturer Received02/21/2024
Supplement Dates FDA Received03/14/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age69 YR
Patient SexMale
Patient Weight108 KG
-
-