Model Number L971210
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/09/2022
Event Type  malfunction  
Event Description
It was reported that when went to pass stem, circulator noticed a puncture hole in top of sterile packaging.Did not occur while opening package or hand off to circulator.No signs of puncture on cardboard box, cardboard insert, or cellophane wrap.Stem never was passed as we noticed prior to passing it.Surgical delay of 12 minutes occurred.Doi: unknown, dor: (b)(6) 2022; affected side: right hip.
 
Manufacturer Narrative
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d3, g1, h3.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this complaint was received for examination.Visual examination found the reported puncture hole in the inner package, however the allegation against the receipt status of the package cannot be confirmed since the outer package has already been opened and no manufacturing nonconformance related to the current complaint was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: a dhr review was performed and 4 non-conformances were found that are not related to the current complaint failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL AMT SZ10 125 STD COL
Type of Device
CORAIL LINE EXTENSION IMPLANTS : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
Manufacturer (Section G)
DEPUY FRANCE SAS 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR   69801
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15481485
MDR Text Key305933777
Report Number1818910-2022-18654
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295455400
UDI-Public10603295455400
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192946
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 09/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberL971210
Device Catalogue NumberL971210
Device Lot Number9056160
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/09/2022
Initial Date FDA Received09/24/2022
Supplement Dates Manufacturer Received09/26/2022
11/01/2022
Supplement Dates FDA Received09/27/2022
11/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexFemale
-
-