It was reported to philips that the patient arrived to the er with ems during cardiac arrest (t3).It was reported that the 1st rhythm check with philips monitor/defib reading rhythm via pads showed ventricular tachycardia.It was reported that the shock delivered successfully.It was reported that during the 2nd rhythm check, the pads were charged to 200j, however a message appeared saying 'pads disconnected, connect cables' even though nothing had changed between rhythm checks.It was reported that new pads were applied but same error message appeared.It was reported that the device was unable to defibrillate the patient despite a shockable rhythm.It was reported that cpr resumed until new monitor/defib obtained, that a successful shock delivered with new monitor but delayed defibrillation.The patient expired.
|
The fse evaluated the device on site.It was determined that this was a malfunction of the therapy cable, which was replaced, and the device was returned to full functionality.The device remains at the customer site and no further evaluation is warranted at this time.
|