Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
It was reported that, the patient experienced torsades de pointes and sustained bradycardia.It was also reported that the right ventricular (rv) lead did not properly capture from insertion and hence, was initially programmed off.Subsequently, the rv lead "fell out," of the cable receptacles and was fractured.The rv lead was repaired and on re-insertion to the patient's cable receptacle and turning the connector knob, the pacemaker paced on the patient's t-wave resulting in torsades de pointes.Cardiopulmonary resuscitation was required.Within the next six weeks, the right atrial (ra) lead failed, where it "fell out," of the cable receptacles and was fractured, and the patient had to rely on rv lead pacing, when required.The leads were disconnected from the cables and the patient was monitored for pacing requirements.Subsequently the patient developed sustained bradycardia and on re-connection of the generator, the pacemaker paced on the t-wave resulting again in torsades de pointes.Cardiopulmonary resuscitation was again required.Approximately two months later, both the rv lead and ra lead were replaced.No further patient complications have been reported as a result of this event.
|