It was reported that, the patient experienced torsades de pointes and sustained bradycardia.It was also reported that the right ventricular (rv) lead did not properly capture from insertion and hence, was initially programmed off.Subsequently, the rv lead "fell out," of the cable receptacles and was fractured.The rv lead was repaired and on re-insertion to the patient's cable receptacle and turning the connector knob, the pacemaker paced on the patient's t-wave resulting in torsades de pointes.Cardiopulmonary resuscitation was required.Within the next six weeks, the right atrial (ra) lead failed, where it "fell out," of the cable receptacles and was fractured, and the patient had to rely on rv lead pacing, when required.The leads were disconnected from the cables and the patient was monitored for pacing requirements.Subsequently the patient developed sustained bradycardia and on re-connection of the generator, the pacemaker paced on the t-wave resulting again in torsades de pointes.Cardiopulmonary resuscitation was again required.Approximately two months later, both the rv lead and ra lead were replaced.No further patient complications have been reported as a result of this event.
|