It was reported that device perforation occurred.The patient presented with an aneurysm in the internal iliac artery.The imager ii angiographic catheter was used for a coiling hypogastric procedure.During the procedure, a 0.035 interlock diamond 6mm coil would not attach.The device was removed 100% intact with traditional pull back.A second interlock coil of the same size was successfully inserted however, upon removal of the imager, it was noted that the coil had perforated through the the sidewall of the catheter at the proximal tip before the angled shape.The procedure was completed with this device.No patient complications were reported.
|