BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGES; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306546
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/07/2022
Event Type  malfunction  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported while using bd posiflush¿ normal saline syringes, mold was found on the syringes.There was no report of patient impact.The following information was provided by the initial reporter: sealed flush has mold.It does seem to have penetrated inside too.The whole box has been quarantined and nothing was dispensed to any patient.
 
Event Description
It was reported while using bd posiflush¿ normal saline syringes, mold was found on the syringes.There was no report of patient impact.The following information was provided by the initial reporter: sealed flush has mold.It does seem to have penetrated inside too.The whole box has been quarantined and nothing was dispensed to any patient.
 
Manufacturer Narrative
Investigation summary: it was reported the syringe had mold and discoloration.To aid in the investigation, two photos were provided for evaluation by our quality team.The photos show two posiflush syringes in sealed packaging flow wraps.Each has a tip cap with discoloration; one is yellowish and the other one is black.No other defects or imperfections were observed.A device history record review was completed for provided material number 306546, lot number 2193626.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.To date, there have been no other similar events reported for this lot.Based on the investigation and with the photo sample analysis the symptom of foreign matter reported by the customer is confirmed, but without the physical sample analysis a probable root cause could not be offered.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGES
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15497247
MDR Text Key306293090
Report Number1911916-2022-00530
Device Sequence Number1
Product Code NGT
UDI-Device Identifier00382903065462
UDI-Public00382903065462
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K003553
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number306546
Device Lot Number2193626
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/07/2022
Initial Date FDA Received09/28/2022
Supplement Dates Manufacturer Received09/28/2022
Supplement Dates FDA Received10/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-