Catalog Number 306546 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
|
Event Description
|
It was reported while using bd posiflush¿ normal saline syringes, mold was found on the syringes.There was no report of patient impact.The following information was provided by the initial reporter: sealed flush has mold.It does seem to have penetrated inside too.The whole box has been quarantined and nothing was dispensed to any patient.
|
|
Event Description
|
It was reported while using bd posiflush¿ normal saline syringes, mold was found on the syringes.There was no report of patient impact.The following information was provided by the initial reporter: sealed flush has mold.It does seem to have penetrated inside too.The whole box has been quarantined and nothing was dispensed to any patient.
|
|
Manufacturer Narrative
|
Investigation summary: it was reported the syringe had mold and discoloration.To aid in the investigation, two photos were provided for evaluation by our quality team.The photos show two posiflush syringes in sealed packaging flow wraps.Each has a tip cap with discoloration; one is yellowish and the other one is black.No other defects or imperfections were observed.A device history record review was completed for provided material number 306546, lot number 2193626.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.To date, there have been no other similar events reported for this lot.Based on the investigation and with the photo sample analysis the symptom of foreign matter reported by the customer is confirmed, but without the physical sample analysis a probable root cause could not be offered.Complaints received for this device and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.H3 other text : see h.10.
|
|
Search Alerts/Recalls
|