• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APIFIX LTD. MID-C 125; POSTERIOR RATCHETING ROD SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APIFIX LTD. MID-C 125; POSTERIOR RATCHETING ROD SYSTEM Back to Search Results
Model Number MID-C 125
Device Problems Unintended Movement (3026); Noise, Audible (3273)
Patient Problem Failure of Implant (1924)
Event Date 08/31/2022
Event Type  malfunction  
Event Description
Patient # (b)(6) index procedure was performed on (b)(6) 2018.On (b)(6) 2022 apifix was notified that patient # (b)(6) is scheduled for a revision surgery on (b)(6) 2022 due to alleged ratchet malfunction.
 
Manufacturer Narrative
Investigation: a review of the device history record confirmed that the device was manufactured and tested according to relevant procedures,and shipped according to manufacturer's specifications.Surgeon, x-ray, information analysis: patient # (b)(6) index procedure was performed on (b)(6) 2018.On (b)(6) 2022 apifix was notified that patient # (b)(6) is scheduled for a revision surgery on (b)(6) 2022 due to alleged ratchet malfunction.On (b)(6) 2022 apifix requested patient x-rays which demonstrate the ratchet malfunction; intermittent noise (no pain) prompted an x-ray at 48 months post-op.Images were provided which show that the ratchet has retracted between 42 months post op to 48 months post op.On (b)(6) 2022 the patient was revised to a new implant.The surgeon maintained the existing polyaxial screws and apifix screw & extender, and only replaced the implant.Corrective action: ratchet malfunction (resulting in a backup of the distraction) can result from physical trauma, practicing high-demand sports, and tissue growth inside the ratchet mechanism.Ratchet malfunction may be a coincidental finding and may also be reported together with pain/curve progression/noise.The company took several actions to mitigate ratchet malfunction: eco-13, the ratchet flat spring component was changed to a thicker spring (from 0.15 to 0.25mm).Eco-59, addition of a stopper pin to prevent the pole from dislocating from the base and inherit from the design to prevent the collapse of the ratchet.In (b)(6) 2020, a highlight of the risk associated with practicing high-demand sports was added to the mid-c training presentation.The present case was (b)(6).The company continues to monitor the effectiveness of the (b)(4) mitigation.Risk assessment: reoperation events are a known risk that was assessed and recorded by the product risk assessment dms-777 rev q this complaint does not change the occurrences rate.The risk of the ratchet malfunction has been assessed and found to be acceptable (dms#777 rev q1 hazard id 1.4 and 1.6).This event does not increase the probability rating.The total rate of ratchet malfunction for any reason is 1.22% the risks have been quantified, characterized, and documented as acceptable within full risk assessment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MID-C 125
Type of Device
POSTERIOR RATCHETING ROD SYSTEM
Manufacturer (Section D)
APIFIX LTD.
1 hacarmel street
kochav yokneam bldg
yokneam elit, 20692 07
IS  2069207
Manufacturer (Section G)
APIFIX LTD.
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692 07
IS   2069207
Manufacturer Contact
alan vaisman
1 hacarmel st.
kochav yokneam bldg
yokneam ellit, 20692-07
IS   2069207
MDR Report Key15507778
MDR Text Key306419109
Report Number3013461531-2022-00048
Device Sequence Number1
Product Code QGP
Combination Product (y/n)N
Reporter Country CodeIS
PMA/PMN Number
H170001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMID-C 125
Device Catalogue NumberAFS-125-050
Device Lot NumberAF-10-01-17
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/31/2022
Initial Date FDA Received09/29/2022
Date Device Manufactured11/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age17 YR
Patient SexFemale
-
-