Investigation: a review of the device history record confirmed that the device was manufactured and tested according to relevant procedures,and shipped according to manufacturer's specifications.Surgeon, x-ray, information analysis: patient # (b)(6) index procedure was performed on (b)(6) 2018.On (b)(6) 2022 apifix was notified that patient # (b)(6) is scheduled for a revision surgery on (b)(6) 2022 due to alleged ratchet malfunction.On (b)(6) 2022 apifix requested patient x-rays which demonstrate the ratchet malfunction; intermittent noise (no pain) prompted an x-ray at 48 months post-op.Images were provided which show that the ratchet has retracted between 42 months post op to 48 months post op.On (b)(6) 2022 the patient was revised to a new implant.The surgeon maintained the existing polyaxial screws and apifix screw & extender, and only replaced the implant.Corrective action: ratchet malfunction (resulting in a backup of the distraction) can result from physical trauma, practicing high-demand sports, and tissue growth inside the ratchet mechanism.Ratchet malfunction may be a coincidental finding and may also be reported together with pain/curve progression/noise.The company took several actions to mitigate ratchet malfunction: eco-13, the ratchet flat spring component was changed to a thicker spring (from 0.15 to 0.25mm).Eco-59, addition of a stopper pin to prevent the pole from dislocating from the base and inherit from the design to prevent the collapse of the ratchet.In (b)(6) 2020, a highlight of the risk associated with practicing high-demand sports was added to the mid-c training presentation.The present case was (b)(6).The company continues to monitor the effectiveness of the (b)(4) mitigation.Risk assessment: reoperation events are a known risk that was assessed and recorded by the product risk assessment dms-777 rev q this complaint does not change the occurrences rate.The risk of the ratchet malfunction has been assessed and found to be acceptable (dms#777 rev q1 hazard id 1.4 and 1.6).This event does not increase the probability rating.The total rate of ratchet malfunction for any reason is 1.22% the risks have been quantified, characterized, and documented as acceptable within full risk assessment.
|