It was reported that during preparation prior to the implant procedure, the device was found to have previously been programmed for use during another previous procedure.The device was not used during the previous procedure due to the device being taken from the shelf by mistake as another type of device was intended for the previous procedure.The device was reprogrammed to original settings.During the implant procedure, sensing anomalies on the discrimination channel and loss of right ventricular sensing were noted on the device.The physician suspected the previous procedure was performed on (b)(6) 2022, however, analysis of the session records by abbott technical support indicated the device was either previously prepared on (b)(6) 2022 or the wrong date was entered on the programmer.The device was explanted and replaced to resolve the event.The patient was in stable condition.
|