• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: QUIDEL CARDIOVASCULAR INC. TRIAGE D-DIMER PANEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

QUIDEL CARDIOVASCULAR INC. TRIAGE D-DIMER PANEL Back to Search Results
Model Number 98100EU
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/22/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation conclusion: the customer's complaint was not replicated during in-house testing of retained devices of lot number t13190rn.Devicesof the complaint lot performed properly when tested with a positive calibrator, no issues with d-dimer recovery were observed.Lot performed within specification.Manufacturing batch records for lot t13190rn were reviewed and found that the lot met final release specifications.Based on the information available, there is no indication of a product deficiency and no correction action is required.
 
Event Description
Customer had discrepant low results with d-dimer tests when compared to the lab results on the same sample.Customer stated that there was a delay in treatment due to these results.The patient was symptomatic prior to testing.No other information provided.
 
Manufacturer Narrative
The final diagnosis was negative for dvt or pe.Clinical outcome matched triage result.
 
Event Description
The final diagnosis was negative for dvt or pe.Clinical outcome matched triage result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAGE D-DIMER PANEL
Type of Device
TRIAGE D-DIMER PANEL
Manufacturer (Section D)
QUIDEL CARDIOVASCULAR INC.
9975 summers ridge road
san diego CA 92121
Manufacturer Contact
chelsea lindgren
9975 summers ridge road
san diego, CA 92121
8583020294
MDR Report Key15521918
MDR Text Key306337879
Report Number3013982035-2022-00015
Device Sequence Number1
Product Code GHH
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K042890
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/11/2023
Device Model Number98100EU
Device Lot NumberT13190RN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/31/2022
Initial Date FDA Received09/30/2022
Supplement Dates Manufacturer Received10/05/2022
Supplement Dates FDA Received10/31/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/28/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
TRIAGE METERPRO PN: (B)(6), SN: (B)(6)
-
-