• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE REV LPS INSRT XSM 14MM; KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE REV LPS INSRT XSM 14MM; KNEE TIBIAL INSERT Back to Search Results
Model Number 1517-60-214
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Joint Dislocation (2374)
Event Date 09/15/2022
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the patient dislocated the insert from the tibia.The main issue was with the patella.Replaced poly and fixed the patella.Doi- (b)(6) 2022, dor- (b)(6) 2022, affected side- left knee.
 
Manufacturer Narrative
Product complaint #: pc-(b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed corrected: g1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE REV LPS INSRT XSM 14MM
Type of Device
KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15526273
MDR Text Key301013323
Report Number1818910-2022-19305
Device Sequence Number1
Product Code KRO
UDI-Device Identifier10603295490975
UDI-Public10603295490975
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1517-60-214
Device Catalogue Number151760214
Device Lot NumberJ63A43
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/15/2022
Initial Date FDA Received10/02/2022
Supplement Dates Manufacturer Received10/25/2022
Supplement Dates FDA Received10/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/18/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN KNEE PATELLA; UNKNOWN KNEE TIBIAL TRAY
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient SexFemale
-
-