Catalog Number 5C4482
Device Problems Fitting Problem (2183); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that a cap would not tighten on the female connector of a minicap extended life pd transfer set; a separation between the female connector and main body of the transfer set was observed and a small blue piece (insert) fell out.This was observed after draining the patient and placing a cap on transfer set.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The event occurred on an unspecified date, over the past weekend from report date.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection with the naked eye on the provided photos noted the female connector separated from the main body.The reported condition was verified.The cause of the condition is related to inadequate solvent application during manufacturing.For the connection issue, this must be performed on the actual device by physically disconnecting the female connector from the patient connector; however, this is not possible due to the lack of a physical sample.Therefore, the reported event could not be verified or refuted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15530555
MDR Text Key303341894
Report Number1416980-2022-05252
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412007731
UDI-Public(01)00085412007731
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152675
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C4482
Device Lot NumberH21I22083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/07/2022
Initial Date FDA Received10/03/2022
Supplement Dates Manufacturer Received10/18/2022
Supplement Dates FDA Received11/08/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/22/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MINICAP
-
-