BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problems
Fitting Problem (2183); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that a cap would not tighten on the female connector of a minicap extended life pd transfer set; a separation between the female connector and main body of the transfer set was observed and a small blue piece (insert) fell out.This was observed after draining the patient and placing a cap on transfer set.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
The event occurred on an unspecified date, over the past weekend from report date.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection with the naked eye on the provided photos noted the female connector separated from the main body.The reported condition was verified.The cause of the condition is related to inadequate solvent application during manufacturing.For the connection issue, this must be performed on the actual device by physically disconnecting the female connector from the patient connector; however, this is not possible due to the lack of a physical sample.Therefore, the reported event could not be verified or refuted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|
|
|