Model Number 37800 |
Device Problem
Electromagnetic Compatibility Problem (2927)
|
Patient Problems
Pain (1994); Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a manufacturer representative regarding a patient (pt) who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that the caller stated pt was having no issues with enterra therapy prior to having a medtronic scs device placed.Once scs device was placed, the pt experienced a buzzing sensation in the enterra location when scs ins was turned on.Pt also notes having shooting pain down their side.The caller states they have limited info on the situation but are going to meet the pt with the scs rep today.The caller stated the devices are on opposites sides of the body.Caller stated the buzzing goes away when scs device turned off.Troubleshooting reviewed considerations, specifically programming of the scs device.
|
|
Manufacturer Narrative
|
H10: review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a manufacturer representative (rep).The rep reported that "this ended up being a non-adverse event.I was there along with the scs rep for one of the pts follow up appointments.The provider turned our device off as well as the scs device.Pt still complained of the pain after the provider acted like she turned the scs device back on along with ours but she never turned either on.Provider said there is nothing wrong with either device and has referred patient for psych evaluation.".
|
|
Search Alerts/Recalls
|
|