MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK ® XS SYSTEM; PROTHROMBIN TIME TEST STRIPS Back to Search Results
Catalog Number 04625374160
Device Problems Loss of Data (2903); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/02/2022
Event Type  malfunction  
Manufacturer Narrative
Occupation is patient/consumer.The product was requested to be returned for investigation however, there are no test strips remaining to return.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.".
 
Event Description
There was an allegation of questionable results from coaguchek xs meter serial number (b)(4) compared to a laboratory result using an unknown method.At 5:35 am the meter result was 6.0 inr.It was alleged that this result was 6.5 inr, not 6.0 inr.The meter memory showed 6.0 inr.At 8:40 am the laboratory result was 4.1 inr.The patient's therapeutic range is 2.5-3.5inr.The patient's interval of testing "depends on result".
 
Manufacturer Narrative
The reporter's meter and test strips were provided for investigation where they were tested using retention controls.Testing results (qc range = 2.4 ¿ 3.6 inr): qc 1: 2.9 inr; qc 2: 2.9 inr; qc 3: 2.9 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The results alleged by the customer were not observed in the meter¿s patient result memory.The obtained qc values were observed in the meter memory without issues.No meter memory issues observed.The customer's allegation could not be reproduced.The returned meter performed according to specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15535537
MDR Text Key306616379
Report Number1823260-2022-03044
Device Sequence Number1
Product Code GJS
UDI-Device Identifier00365702127104
UDI-Public00365702127104
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/31/2023
Device Catalogue Number04625374160
Device Lot Number58526222
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/06/2022
Initial Date FDA Received10/04/2022
Supplement Dates Manufacturer Received10/18/2022
Supplement Dates FDA Received11/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ACTEMRA INJECTION; ATORVASTATIN; FLONASE; LEVOTHYROXINE; LISINOPRIL; MAGNESIUM; PREDNISONE; SOTALOL; TAMSULOSIN; WARFARIN
Patient Age62 YR
Patient SexMale
-
-