• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE CORI; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10024
Device Problem Computer Operating System Problem (2898)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/10/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, during machine set up for cadaver workshop, the real intelligence cori was not getting booted.There was a continuous beep sound from the machine.They restarted the machine a couple of times, still the issue was not resolved.As this happened in a bioskills laboratory, there was not patient involvement.
 
Manufacturer Narrative
Section h3, h6: the real intelligence cori, part number rob10024, serial number (b)(6), intended for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.While all products meet required manufacturing specifications prior to release a serial number, lot number, part revision or software version is required to link the device to a dhr or nc investigation.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with console software bug.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE CORI
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15535781
MDR Text Key302797563
Report Number3010266064-2022-00619
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757420
UDI-Public00885556757420
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K193120
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10024
Device Catalogue NumberROB10024
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/10/2022
Initial Date FDA Received10/04/2022
Supplement Dates Manufacturer Received11/06/2022
Supplement Dates FDA Received11/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-