AIZU OLYMPUS CO., LTD. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/05/2022
Event Type  malfunction  
Manufacturer Narrative
An olympus field service engineer (fse) was dispatched to evaluate and repair the device.The fse observed, the yellow connector was chipped on the outer ring, center pin was intact and there was no disruption of fluid flow through the channel.Fse replaced the yellow connector to resolve the reported issue.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus technical assistance center (tac), the endoscope reprocessor had a broken yellow connector.There was no harm or user injury reported due to the event.
 
Event Description
There was no patient care or procedure issue.The facility had two reproccessors, and was able to label the affected reproccessor as "not for use/dated".
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Section b5 was updated with additional information about the event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The cause of the reported event cannot be conclusively determined.Based on the results of the investigation, it is likely the connector could not be connected as some impact was added to the connector and the connector became loose and came apart.Probable causes for the event include: "* accumulated stress over time which caused the connector to get loose * unintentional impact to the connector with something hard." per the instructions for use: ¿check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents.¿ and ¿warning -do not use the equipment if any connector seems to be damaged or defective.Using the equipment when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral device or facilities near the equipment.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15538606
MDR Text Key306681205
Report Number9610595-2022-02562
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/06/2022
Initial Date FDA Received10/04/2022
Supplement Dates Manufacturer Received10/05/2022
Supplement Dates FDA Received11/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-