Model Number 866424 |
Device Problems
Misconnection (1399); Insufficient Information (3190)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A follow-up report will be submitted upon completion of the investigation.
|
|
Event Description
|
The customer reported that they cannot hear the alarm from the unit.The device was in use at time of event, there was no adverse event reported.
|
|
Event Description
|
The customer reported that the pic ix c st efp indicating that the customer could not hear the alarm.There was no patient or user harm.The philips field service engineer (fse) went onsite to assist the customer.The fse found that the network cable was plugged into the wrong port and tha moved the network cable to a different port.After moving the cable, the sound was received remotely to the central station.
|
|