• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC PIC IX C ST EFP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC PIC IX C ST EFP Back to Search Results
Model Number 866424
Device Problems Misconnection (1399); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2022
Event Type  malfunction  
Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.
 
Event Description
The customer reported that they cannot hear the alarm from the unit.The device was in use at time of event, there was no adverse event reported.
 
Event Description
The customer reported that the pic ix c st efp indicating that the customer could not hear the alarm.There was no patient or user harm.The philips field service engineer (fse) went onsite to assist the customer.The fse found that the network cable was plugged into the wrong port and tha moved the network cable to a different port.After moving the cable, the sound was received remotely to the central station.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIC IX C ST EFP
Type of Device
PIC IX C ST EFP
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
3000 minuteman road
andover MA 01810
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
3000 minuteman road
andover MA 01810
Manufacturer Contact
jeanne ahearn
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key15542283
MDR Text Key302420244
Report Number1218950-2022-00902
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866424
Device Catalogue Number866424
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/13/2022
Initial Date FDA Received10/05/2022
Supplement Dates Manufacturer Received01/06/2023
Supplement Dates FDA Received01/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-