SMITH & NEPHEW, INC. JRN II BCS LCK FEM IMPLANT IMPACT; PROSTHESIS,KNEE,PATELLOFEMOROTIBIAL,SEMICONSTRAINED,CEMENTED,POLYMERMETALPOLYMER
|
Back to Search Results |
|
Model Number 74012812 |
Device Problem
Flaked (1246)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, during surgery, a jrn ii bcs lck fem implant impact was wear after several use.Device outside the patient.The procedure was completed without delay using a s+n back-up device.Patient was not harmed.
|
|
Manufacturer Narrative
|
The associated device was returned and evaluated.A visual inspection of the returned device confirms a large piece of the cam arm broke off the device.The broken piece was not returned.The device shows signs of significant wear and use.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|