• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE ENVELLA BED; BED, AIR FLUIDIZED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE ENVELLA BED; BED, AIR FLUIDIZED Back to Search Results
Model Number P0819A
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/29/2022
Event Type  malfunction  
Event Description
Hillrom received a report from a hillrom technician stating the power cord burnt out.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as (b)(4).
 
Manufacturer Narrative
The hillrom technician found the power cord needed to be replaced.Per the hillrom service manual, perform annual preventive maintenance procedures to make sure all bed components are functioning as originally designed.The user manual instructs that power cords require inspection routinely to ensure no damage has occurred during use.The manuals also include warnings such as incorrect use or handling may result in damage to the power cord.They also instruct that if damage has occurred remove immediately, and to properly remove the power cord from the electrical outlet during transfer or movement of the device.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in may 6, 2022.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the power cord to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVELLA BED
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
melrita gilliam
1069 state route 46 east
batesville, IN 47006
3127289851
MDR Report Key15546355
MDR Text Key301855095
Report Number1824206-2022-00440
Device Sequence Number1
Product Code INX
UDI-Device Identifier00887761013834
UDI-Public010088776101383411180117
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP0819A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/29/2022
Initial Date FDA Received10/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-