MEDTRONIC NAVIGATION, INC STEALTHSTATION FLEXENT; EAR, NOSE, AND THROAT STEREOTAXIC INSTRUMENT
|
Back to Search Results |
|
Model Number 9736242 |
Device Problems
Electromagnetic Interference (1194); Imprecision (1307); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/23/2022 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding a navigation device used during an endoscopic skull base procedure.It was reported that in the last 12 months, the surgeon encountered a "target registration error (tre)" ten times.It was reported that each occurrence happened intraoperatively.In each occurrence, intraoperative re-registration was required due to a navigation inaccuracy.Of those ten occurrences, there were eight times that the surgeon alleged a "tre greater than 2.0 millimeters".There was no reported impact on the procedure and no adverse events were encountered. it was reported that there was there was one occurrence of metallic interference with the electromagnetic (em) navigation with no clinical or patient impact.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|