• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION; DEFIBRILLATION LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION; DEFIBRILLATION LEAD Back to Search Results
Model Number 7122/65
Device Problems Failure to Capture (1081); High impedance (1291); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2022
Event Type  Injury  
Event Description
It was reported during in clinic follow up that the right ventricular lead exhibited high pacing impedance, high defibrillation impedance, loss of capture, and loss of sensing.Diagnostic imaging did not reveal any physical anomalies.The lead was explanted and successfully replaced.The patient was stable and asymptomatic.
 
Manufacturer Narrative
The reported events of high pacing lead impedance, loss of capture, high defibrillation impedance and loss of sensing were not confirmed.As received, a complete lead was returned in one piece.Electrical tests and x-ray examination did not find any indication of conductor fractures or internal shorts.Visual inspection of the lead did not find any anomalies.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION
Type of Device
DEFIBRILLATION LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15567242
MDR Text Key301471520
Report Number2017865-2022-39792
Device Sequence Number1
Product Code LWS
UDI-Device Identifier05414734502153
UDI-Public05414734502153
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P950022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model Number7122/65
Device Catalogue Number7122-65
Device Lot NumberA000109811
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/21/2022
Initial Date FDA Received10/07/2022
Supplement Dates Manufacturer Received11/16/2022
Supplement Dates FDA Received11/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-