Model Number KNEE SCORPION
Device Problems Break (1069); Misfire (2532)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2022
Event Type  malfunction  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
On (b)(6) 2022 it was reported by a sales representative via sems that an ar-12990 knee scorpion misfired multiple times and the tip broke.No patient affect.Additional information has been requested.
 
Manufacturer Narrative
Complaint confirmed.Visual evaluation noticed that the jaw is broken off.This is probably caused by applying excessive force while grasping and manipulating tissue, or when applying excessive leveraging forces.No broken parts were returned for inspection.Functional testing not performed due to the damage of the device.The observed condition is attributed to undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNEE SCORPION
Type of Device
ACCESSORIES, ARTHROSCOPIC
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15571461
MDR Text Key306690576
Report Number1220246-2022-05598
Device Sequence Number1
Product Code NBH
UDI-Device Identifier00888867196322
UDI-Public00888867196322
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 12/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKNEE SCORPION
Device Catalogue NumberAR-12990
Device Lot Number10253206
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/22/2022
Initial Date FDA Received10/10/2022
Supplement Dates Manufacturer Received09/22/2022
Supplement Dates FDA Received12/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2018
Is the Device Single Use? No
Type of Device Usage Initial