Model Number 33 +4 LAT/24 GLENOSPHERE
Device Problems Fracture (1260); Loose or Intermittent Connection (1371)
Patient Problem Pain (1994)
Event Date 01/22/2021
Event Type  Injury  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
On (b)(6) 2022, it was reported by a patient via email that she underwent a reverse shoulder replacement on (b)(6) 2021, and is now experiencing sudden pain to her right shoulder when raising her arm overhead.X-rays suggest a stress fatigue fracture through the inferior screw.This demonstrates loosening of the left glenoshpere/baseplate.No additional information provided.Additional information requested.Additional information provided on (b)(6) 2021: implanted devices are as follows: right reverse tsr, arthrex univers 6mm stem with 36/+2 right suture cup, small 33/+3 humeral insert, 24mm mono-block base plate with 3 locking screws,+4/33mm glenophere.A ct scan of the right shoulder obtained at lehigh valley hospital is available for review.This demonstrates appropriate positioning of reverse tsr components.A minimal amount of fluid is seen.There is degenerative change to the ac joint.No lucency around the implants.Due to the artifact the screw fracture is not visible.At this time a revision surgery will not take place.Additional information requested.Additional information provided 09/21/2022: implanted devices are as follows: ar-9560-24p, batch 8164; ar-9502f-36rcpc, batch: 19.04738; ar-9564-2433-lat, batch: 19.00839; ar-9503-3633-3, batch: 19.02941; ar-9563-20, batch: 2019004605; ar-9563-32, batch: 2019004423; ar-9562-32nl, batch: 2019002834; ar-9501-06p, batch: 19.04248.It was reported by the patient that the revision took place on (b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
33 +4 LAT/24 GLENOSPHERE
Type of Device
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15571915
MDR Text Key301504298
Report Number1220246-2022-05583
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888867295650
UDI-Public00888867295650
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173900
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number33 +4 LAT/24 GLENOSPHERE
Device Catalogue NumberAR-9564-2433-LAT
Device Lot Number19.00839
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/21/2022
Initial Date FDA Received10/10/2022
Date Device Manufactured10/01/2019
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-