This report is being supplemented to correct information provided in the initial report, and to provide additional information based on the legal manufacturer's final investigation.The date received by manufacturer was sept.15, 2022.The date returned to manufacturer was sept.21, 2022.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.Based on the results of the investigation, the broken fibers caused the blurry image.The root cause of the broken fibers was not able to be determined.It was determined the user could have completed the procedure if they had prepared a spare device, as per the device's instructions for use (ifu): "spare equipment - be sure to prepare another endoscope to avoid interruption of the examination due to equipment failure or malfunction.¿ the suggested event can be detected by handling the device in accordance with the information provided in the device's ifu under "inspection of the endoscopic system." olympus will continue to monitor field performance for this device.
|