• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 25MM CENTRAL SCREW, MODULAR; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 25MM CENTRAL SCREW, MODULAR; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number 25MM CENTRAL SCREW, MODULAR
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2022
Event Type  malfunction  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
It was reported on 09/22/2022 by a sales representative via email that an ar-9560-28-2 baseplate and ar-9561-25s modular screw would not compress together with the morse taper using the taper assembly press.This was discovered during a case with no patient harm.Per facility: " it was confirmed that the compression line between maximum and minimum was achieved but up insterting the implant the two pieces became disassembled.We were able to retrieve both parts safely without harming the patient.We then assembled two new parts, same sizes, and in the same fashion without issues.".
 
Manufacturer Narrative
Complaint is confirmed.One unpackaged ar-9561-25s, 25mm central screw, modular serial/batch number (b)(6), was received for investigation.Functional testing found that the device does not attach to the returned ar-9560-28-2 - 28mm +2 lat baseplate, modular.Visual evaluation found that the central screw taper connector hex was damaged, losing its geometry form.Complaint use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
25MM CENTRAL SCREW, MODULAR
Type of Device
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key15576790
MDR Text Key306243858
Report Number1220246-2022-05600
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888867296565
UDI-Public00888867296565
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K173900
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25MM CENTRAL SCREW, MODULAR
Device Catalogue NumberAR-9561-25S
Device Lot Number14952286
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/23/2022
Initial Date FDA Received10/11/2022
Supplement Dates Manufacturer Received09/23/2022
Supplement Dates FDA Received02/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/20/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-