A physician reported while pulling out the cutter from the eye, the trocar came out with the cutter.Upon re-inserting the trocar and attempting to insert the cutter, it was noticed that the tip of the cutter was missing.The tip of cutter was found on the table.No foreign material was observed in the eye at this time during cataract/vitrectomy combination procedure.The outer cylinder of the cutter was on the table, but the doctor was concerned about the inner cutter.There was no patient harm reported.
|
Additional information provided in d.9., h.3., h.6.And h.10.One opened probe was received, with a tip protector, in a tray.The sample was visually inspected and found to be nonconforming with the probe needle broken at the port and orange/brownish foreign material on the remaining are of the port.No functional testing with ring gage could be performed due to needle port broken condition.The probe was disassembled and the components inspected.Nominal wear observed on inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Wear marks observed at one location on the inner cutter.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification (rfid) tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The complaint evaluation was unable to confirm the report product fit issue due to the needle port broken condition, which could be a contributing factor of the trocar coming out with the cutter at the time the event was reported.The complaint evaluation confirms the probe port tip was broken.The exact root cause of the broken tip cannot be determined from this evaluation.An internal investigation has been completed and action has been implemented to reduce the frequency of broken tips.No additional actions has have been taken by the manufacturing as the reported product fit issue was unable to be confirmed.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.All assembled probe needle diameters are also 100% tested for fit into a ring gauge to insure the probe needle does not exceed a trocar opening.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is:(b)(4).
|