Model Number M00522610
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2022
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic polypectomy procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip could not be deployed and fell off from the biopsy hole during withdrawal of the scope.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
Block h6: medical device code a15 captures the reportable event of clip could not be deployed.Block h10: investigation analysis: the returned resolution clip device was analyzed, and a visual evaluation noted that the device was returned without the clip assembly attached.The clip assembly was returned in a separate bag.Microscopic examination was performed, and it was found that the clip had evidence of both activations performed.Additionally, the yoke was returned detached from the control wire, which is evidence of a proper deployment, and the locking tabs were opened due to normal deployment.No problems with the device were noted.The reported event of clip could not be deployed was not confirmed.The returned device review showed no evidence of either the alleged(s) or any defect which could have contributed to the event.Therefore, the most probable root cause for this complaint is no problem detected.
 
Event Description
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic polypectomy procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip could not be deployed and fell off from the biopsy hole during withdrawal of the scope.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION CLIP
Type of Device
CLIP, HEMOSTATIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15583112
MDR Text Key306510975
Report Number3005099803-2022-05802
Device Sequence Number1
Product Code MCH
UDI-Device Identifier08714729504795
UDI-Public08714729504795
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K040148
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/02/2022
Device Model NumberM00522610
Device Catalogue Number2261
Device Lot Number0024952676
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/15/2022
Initial Date FDA Received10/11/2022
Supplement Dates Manufacturer Received11/02/2022
Supplement Dates FDA Received11/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexMale
Patient Weight77 KG
-
-