• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER INFINITI VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750469
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/28/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that tubing of an ophthalmic phacoemulsification (phaco) handpiece disconnected during a cataract surgery.The surgery was completed with no patient harm.This is the second of three reports for the reported event from this facility.
 
Manufacturer Narrative
The lot/batch/serial is unknown.Therefore, a service history review cannot be performed.Based on the information available, the customer reported event cannot be confirmed.Specific product identifiers (lot number, batch number, and/or serial number) were not provided and could not be determined at this time.However, all device history records (dhr) are reviewed prior to product release to ensure the product was manufactured in compliance with the device master record and meets release criteria.A review for complaints reported against this lot/batch/serial number cannot be performed as the lot/batch/serial number is unknown.Based on the information obtained, the root cause of the reported event is inconclusive.The manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITI VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15585581
MDR Text Key306820780
Report Number2028159-2022-01444
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657504695
UDI-Public00380657504695
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750469
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/14/2022
Initial Date FDA Received10/12/2022
Supplement Dates Manufacturer Received11/02/2022
Supplement Dates FDA Received11/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM
-
-