Model Number 3100 A
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/16/2022
Event Type  Injury  
Manufacturer Narrative
Vyaire field service technician went onsite and evaluated the device.Found circuit calibration adjustment maxed out.Technician adjusted correctly and performed circuit calibration and performance checkout with normal expected results.Ran device at patient settings with normal expected results.Technician talked to staff and figured out that sound they were hearing was the pressure limit adjustment coming into play.Reviewed basic setup and operation of this control but referred them back to dept policy for actual adjustment of this control.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
It was reported to vyaire medical that rumbling sound was heard when pressure was set to 8 during use on a patient on the 3100 a ventilator.The customer reported the patient was removed from the device.The customer reported there was no patient harm associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL INC.
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key15589416
MDR Text Key301626830
Report Number2021710-2022-16718
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20220316
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/16/2022
Initial Date FDA Received10/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-