• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problems Break (1069); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2022
Event Type  malfunction  
Event Description
A user facility returned the olympus, model clv-190, evis exera iii xenon light source to olympus for maintenance due to a report of dust and debris in the unit.Upon inspection and testing of the returned device, a "b30" error occurred and multiple screws were found broken and/or missing and stuck in the casing.This report is submitted due to the findings of a "b30" error and multiple screws broken and/or missing and stuck in the casing.As the problem was found during in-house service of the device, there was no patient involvement.
 
Manufacturer Narrative
The device was evaluated by olympus.A "b30" scope communication error occurred due to poor contact at the scope connector.In addition, multiple screws were noted missing and broken; parts of the screws were found in the casing.The casing chassis was noted corroded.The investigation is ongoing and the definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the b30 error and broken screw being stuck in the casing could not be identified.However, it¿s likely the cause of the b30 error is related to part of the broken screw being stuck in the casing.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15590600
MDR Text Key306615836
Report Number3002808148-2022-03073
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/14/2022
Initial Date FDA Received10/12/2022
Supplement Dates Manufacturer Received11/02/2022
Supplement Dates FDA Received11/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-