• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION VERCISE GEVIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION VERCISE GEVIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-1200-S
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pocket Erosion (2013)
Event Date 08/16/2022
Event Type  Injury  
Event Description
It was reported by the patients mother that the patients deep brain stimulator (dbs) implantable pulse generator (ipg) was starting to erode out of her skin.The patient was seen by the neurosurgery team and given topical antibiotics and the implant site was bandaged.The incision site has healed up to the point where there is no exposure of the ipg through the skin and the patient is receiving adequate therapy.After review of the ipg battery charge, it was found that the patients ipg had been completely drained throughout the month of (b)(6) 2022 to the point that no dots were visible.The patient began charging again at the start of (b)(6) 2022 and had charged several times to get up ipg back up to normal range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE GEVIA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key15592367
MDR Text Key301625504
Report Number3006630150-2022-05481
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729984443
UDI-Public08714729984443
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/14/2022
Device Model NumberDB-1200-S
Device Catalogue NumberDB-1200-S
Device Lot Number741689
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/22/2022
Initial Date FDA Received10/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age14 YR
Patient SexFemale
-
-