• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TUBES BLUE LINE CLASSIC; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX TUBES BLUE LINE CLASSIC; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 100/515/080
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2018
Event Type  malfunction  
Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms#: 617147.No product information has been provided to date.There were no non-conforming reports (ncrs) initiated for the lot or anomalies identified in the device history record (dhr) review.A product sample was received for evaluation without its original packaging and in used condition.Visual and functional testing were performed.Visual inspection was performed at a distance of 12 inches under normal lighting to received unit, in order to detect any damage on the cuff or airline.During the visual inspection there were no discrepancies were found.During the functional testing the samples was leak tested under water and during the test it was found the cuff was leaking.The root cause was unable to be determined.As preventive action, manufacturing personnel were notified by the supervisor.
 
Event Description
It was reported that during the pre-use check it was discovered that the device leaked out air no patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TUBES BLUE LINE CLASSIC
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15596127
MDR Text Key306793257
Report Number3012307300-2022-23755
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K030570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/08/2023
Device Catalogue Number100/515/080
Device Lot Number3565173
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/21/2022
Initial Date FDA Received10/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-