SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 7-8 9MM RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/M
|
Back to Search Results |
|
Model Number 74035271 |
Device Problem
Flaked (1246)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/20/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that, during tka, a fragment of insert of a jrny ii bcs art isrt trl sz 7-8 9mm rt broke away when removing from base plate.It is unknown how surgery was completed.The procedure was completed, without delay.Patient was not harmed.
|
|
Manufacturer Narrative
|
(b)(4).
|
|
Manufacturer Narrative
|
Additional information: d4, h4 h3, h6: the associated device was returned and evaluated.A visual inspection of the returned trial confirms pieces of the device have broken off.These pieces were not returned.This device exhibits signs of significant wear and use.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it cannot longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|
|