Model Number 16026545
Device Problem Malposition of Device (2616)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2022
Event Type  Injury  
Event Description
It was reported that the patient underwent an initial posterior fixation procedure from t10 to l1.Subsequently, it was identified through postoperative ct imaging that the right pedicle screw at the t12 level was malpositioned medially.The patient did not report experiencing any pain or other adverse symptoms as a result of the placement.Approximately seven (7) days later, the patient underwent a revision procedure to remove the malpositioned screw.No new screw was implanted into the right t12 location.No further patient impact was reported.No additional information was available.
 
Manufacturer Narrative
It is unknown which screw lot number was associated with the reported event: n314884 or n328363.Manufacture date: 06 feb 2022 or 13 jun 2022.The reported event was unable to be confirmed due to limited information received from the customer.No device was returned to nuvasive for evaluation; further, no operative notes and/or radiograph images were provided for review of usage/technique.A review of the device history records was performed and no discrepancies relevant to the reported event were found.Based on the information provided by the complainant, the root cause is considered to be an inadvertent use error and the result of screw malplacement during the implantation procedure.Labeling review: "potential adverse events and complications: as with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries include: early or late infection; damage to blood vessels, spinal cord or peripheral nerves; pulmonary emboli; loss of sensory and/or motor function; impotence; and permanent pain and/or deformity." "warnings, cautions and precautions: the safety and effectiveness of these devices for any other conditions are unknown.The implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." "pre-operative warnings: care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." if any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE RELINE SYSTEM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INC.
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key15611115
MDR Text Key301821336
Report Number2031966-2022-00205
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00887517556639
UDI-Public887517556639
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K143684
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number16026545
Device Catalogue NumberN/A
Device Lot NumberSEE H10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/21/2022
Initial Date FDA Received10/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient SexPrefer Not To Disclose
-
-