Concomitant medical product ts: product id: 4351-35, serial#: (b)(4), implanted: (b)(6) 2021, explanted: (b)(6) 2022, product type: lead.Product id 4351-35, serial#: (b)(4), implanted: (b)(6) 2021, explanted: (b)(6) 2022, product type: lead.Other relevant device(s) are: product id: 4351-35, serial/lot #: (b)(4), ubd: 08-aug-2023, udi#: (b)(4).Product id: 4351-35, serial/lot #: (b)(4), ubd: 08-aug-2023, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
Information was received from a healthcare provider (hcp) regarding a patient who was implanted with an implantable neurostimulator (ins) for gastric stimulation.It was reported that patient was scheduled for a lead replacement today because of suspected lead migration in stomach wall.During the lead replacement, the surgical technician did not fully loosen the front header blocker screws and apparently applied strong force on the leads when removing, so much so that it moved the top upper set screw (electrode 0) from its normal seating preventing it from being tightened.Because of this, the site used a new stimulator and was able to connect, check impedance (normal) and program.Patient is doing fine.The source of this issue was from the operation room tech not fully loosening set screw and applying strong pressure in removing leads.The issue was resolved with replacement.
|