H4: the lot was manufactured from september 29, 2021 - september 30, 2021.H10: one (1) actual device was received for evaluation.The other sample was not received and therefore, could not be evaluated.Visual inspection was performed which observed a damaged section at the middle area of the silicone tubing.A visual inspection of the area under magnification revealed a tear/hole approximately 0.50 inches in length.Functional testing was performed which revealed a leak in the affected area and not at the blue connector area.The reported condition of leak was verified.The cause of the condition could not be determined; however, a possible cause of the damage was from the repeater pump rotor impact.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|