Model Number BEA25-80/I16-40
Device Problems Collapse (1099); Insufficient Information (3190)
Patient Problems Aneurysm (1708); Failure of Implant (1924)
Event Date 09/28/2022
Event Type  Injury  
Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Event Description
The patient was initially implanted with a bifurcated stent graft and a suprarenal aortic extension to treat an abdominal aortic aneurysm (aaa).Approximately two (2) years post initial procedure an indeterminate origin endoleak was identified.The physician elected to reline the original implanted devices with another afx2 bifurcated stent graft to resolve this event.The patient was reported as doing fine post secondary procedure and was discharged.
 
Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows that the indeterminate endoleak was found to be multiple type 2 endoleaks (lumbars).The additional endovascular procedure is confirmed.This is moderately consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest buckling of the proximal right common iliac stent and sac growth of 5mm occurred that were not included in the event as reported.These findings were discovered during a review of the ct scan dated 03/02/2021 and the angiogram dated 10/13/2022.The event is most likely anatomy related.It was reported that a "cutdown procedure was performed recently that damaged the right main body limb".It is unclear if this contributed to the reported event.No procedure related harms for this complaint were identified.The final patient status was reported to be doing fine.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.
 
Event Description
The patient was initially implanted with a bifurcated stent graft and a suprarenal aortic extension to treat an abdominal aortic aneurysm (aaa).Approximately two (2) years post initial procedure an indeterminate origin endoleak was identified.The physician elected to reline the original implanted devices with another afx2 bifurcated stent graft to resolve this event.The patient was reported as doing fine post secondary procedure and was discharged.Additional information: an internal clinical assessment determined buckling of the proximal right common iliac stent and sac growth of 5mm occurred that were not included in the event as reported.These findings were discovered during a review of the computed tomography (ct) scan dated 03/02/2021 and the angiogram dated 10/13/2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Type of Device
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key15629479
MDR Text Key301978962
Report Number2031527-2022-00252
Device Sequence Number1
Product Code MIH
UDI-Device Identifier00818009014733
UDI-Public(01)00818009014733(17)230624
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/24/2023
Device Model NumberBEA25-80/I16-40
Device Lot Number2365169010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/28/2022
Initial Date FDA Received10/18/2022
Supplement Dates Manufacturer Received11/28/2022
Supplement Dates FDA Received12/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/15/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AFX SUPRARENAL AORTIC EXTENSION, LOT # 1818592R004
Patient Outcome(s) Required Intervention;
Patient Age84 YR
Patient SexMale
-
-