Model Number LTF-S190-10 |
Device Problems
Image Display Error/Artifact (1304); Poor Quality Image (1408)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation confirmed, the noise in the image during angling of the device was due to a short-circuit of the image sensor cable.The issue of the abnormal color tone was due to a damaged charge-coupled device.The bending section cover had scratches and was missing adhesive.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental will be submitted on completion of investigation or if any additional information is available.
|
|
Event Description
|
The customer reported to olympus, the video scope produced a bad image.The color tone of the video scope image was abnormal and the image was noisy with vertical lines.The device was returned, and it was confirmed that the image had magenta color tone and there was noise in the image during angulation in the up/down direction.There were no reports of patient harm associated with the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The event date and additional information about the event is also captured in this report.A review of the device history record (dhr) found no deviations that could have caused or contributed to the image failure.The dhr confirmed that the subject device was shipped in accordance with the specifications.A definitive root cause for this issue was not established.However, it is probable that the issue occurred because of a defective image sensor unit.The cause of the defect was not determined.Olympus will continue to monitor field performance for this device.
|
|
Event Description
|
It was reported that the issue was noticed during pre-inspection for use for a procedure.
|
|