• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-10
Device Problems Image Display Error/Artifact (1304); Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/13/2022
Event Type  malfunction  
Manufacturer Narrative
A device evaluation confirmed, the noise in the image during angling of the device was due to a short-circuit of the image sensor cable.The issue of the abnormal color tone was due to a damaged charge-coupled device.The bending section cover had scratches and was missing adhesive.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental will be submitted on completion of investigation or if any additional information is available.
 
Event Description
The customer reported to olympus, the video scope produced a bad image.The color tone of the video scope image was abnormal and the image was noisy with vertical lines.The device was returned, and it was confirmed that the image had magenta color tone and there was noise in the image during angulation in the up/down direction.There were no reports of patient harm associated with the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The event date and additional information about the event is also captured in this report.A review of the device history record (dhr) found no deviations that could have caused or contributed to the image failure.The dhr confirmed that the subject device was shipped in accordance with the specifications.A definitive root cause for this issue was not established.However, it is probable that the issue occurred because of a defective image sensor unit.The cause of the defect was not determined.Olympus will continue to monitor field performance for this device.
 
Event Description
It was reported that the issue was noticed during pre-inspection for use for a procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15631779
MDR Text Key302000870
Report Number9610595-2022-03074
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170310386
UDI-Public04953170310386
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2022
Initial Date FDA Received10/19/2022
Supplement Dates Manufacturer Received10/20/2022
Supplement Dates FDA Received11/04/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/20/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-